Calibration & the problems it poses
A Word About Winkler...Winkler remains the "gold" standard of calibration techniques. Winkler is based on chemistry, which means, as long as your reagents are fresh and prepared properly and accurately, your calibration will be accurate. We DO see differences when Winkler is compared to Probe technology...but the differences are in absolute DO measurements; critical QC results are the same. Winkler was, at one point, THE best available calibration technique. The reality is, however, that advances in probe technology combined with the simple fact that Winkler calibration is time and labor intensive, are slowly contributing to Winkler's demise. Certainly, if Winkler works for you, and you have the time required, by all means continue to use it. Alternatively, rest assured that you CAN obtain consistent and reliable calibrations using a DO probe. The keys to successful Winkler calibration are:
Apples to Apples? or Apples to Oranges? With air-saturated WATER calibration, you are treating samples and calibration alike...
the probe in both cases is submersed in sample (or dilution water).
With water-saturated AIR calibration, samples and calibration are treated differently...the probe is in air for calibration, but submersed for sample analysis.
For many years, we have referred to AIR calibration as a case of "apples to oranges" since the probe sits in air for calibration but is immersed in sample (or dilution water) during sample analysis. This is distinctly different from the WATER (saturated air) calibration in which the probe is always immersed in sample (or dilution water). Historically, this difference has been the root cause of many a blank failure. At the heart of this issue is physics. We know we can easily saturate water with oxygen by simply shaking the water in an environment that offers sufficient exposure to oxygen in the air. This is best accomplished using a large container which is half full of dilution water, offering sufficient headspace for air (and thus oxygen). Maintaining a water-saturated air environment, however, is a far more difficult proposition. Consider an air calibration performed in the winter. It's quite possible that the atmosphere in a BOD bottle containing a little water, and sealed with the probe, represents 100% saturation (100% relative humidity). This is usually evidenced by the formation of condensation droplets on the probe tip. What's the first thing you do? Remove the probe from the bottle and either shake or wipe off the probe tip to remove the drops...right? Of course...but what have you done to the water saturated air in the bottle? If your lab is like most, the room air likely has a humidity of about 50% (or less). Removing the probe from the bottle causes a draft effect, which pulls the water-saturated air out of the bottle...to be replaced by air of much lower relative humidity. Now comes the physics. So you seal up the bottle again by replacing the probe. Physics pop quiz time: How longs does it take for the air in the bottle to become saturated with waster again. Here's a tip...it doesn't occur in just a few minutes. Frankly, this "passive saturation" can take hours, depending on barometric pressure, the room temperature, the amount of water in the bottle, and the amount of headspace of the bottle. Bottom line: There are just too many variables, and that's exactly why your blanks can deplete 0.3 mg/L one day and "create" 0.3 mg/L the next day. That's why we haven't been huge fans of AIR calibration. But...things are changing; technology is improving. WATER vs. AIR calibration using the same sample
WATER Calibration example.
AIR Calibration example. Chemists at the State Lab of Hygiene performed AIR and WATER calibrations in succession with the same sample, DO meter and probe. They noted that the AIR calibration result was higher by 0.15 mg/L (8.68 vs. 8.53) . NOTE that 8.68 divided by 1.023 equals 8.48, which represents the true saturation of oxygen at 742 mm pressure and 22.2° C. The factor 1.023 (102.3%) represents a correction factor used to adjust results obtained following AIR calibration to account for difference in temperature equilibration time. The calibration for new technology (dual thermistor) probes is 102.3% and 101.7% for older polarographic probes. NOTE: These observations come specifically from one specific DO meter manufacturer and model number. Bottom line: the value shown after air calibration will be high but if immediately placed in water will read correctly.
Common Sources of Error in the Water Saturated AIR Calibration
It takes time for the probe equilibrate initially after removing the droplets from the tip. It also takes longer to equilibrate in air than water! That’s why some have more success with the air-saturated WATER approach. Key to successful water saturated air calibration
Your calibrations will likely work even if you don’t wait for the air to be 100% saturated with water as long as you do your calibration the same EVERYDAY! One operator’s SOP for consistency:
Source: Joe Flannigan, Blanchardville Wastewater Treatment Plant; former Lab-of-the-Year Winner. Conclusions: AIR vs. WATER Calibration
Correct Calibration ProtocolsWater-Saturated AIR
Air-Saturated WATER
Summary of Calibration Techniques
How do you KNOW DO measurements are accurate?The answer is: You need a "known" standard. If you have a standard which you know to be a specified dissolved oxygen concentration, and you measure that standard and obtain a reading equal to (or acceptably close to) that true value, then you have documentation to support the ability to obtain accurate measurements. We also understand that labs have a lot of QC responsibilities already, so the best part of all this is that you can analyze a known standard without having to add any additional analyses! That's because, the initial DO reading of your calibration blank IS a known standard. The process can be summarized as:
Remember: Temperature ± 0.5 °C translates to ± 0.1 mg/L DO Remember: Pressure ± 5 mm translates to ± 0.06 mg/L DO Last Revised: Wednesday January 03 2007
|