Troubleshooting BOD Testing
|
Is it: Contamination? Calibration? Or Seed?
Contamination?
- Can affect BOTH blanks and GGA
- Tends to be LARGE effect
- Tends to be HIGH bias (GGA high)
- Must determine whether the contamination is "food" or "bugs"
Calibration?
- Mainly affects blanks
- Tends to be SMALL effect
- Can be LOW or HIGH bias (blanks deplete > 0.2 mg/L or GAIN > 0.2 mg/L)
Or Seed?
- Mainly affects GGA
- Typically an issue of not enough seed
- Tends to be LARGE effect
- Tends to be LOW bias (GGA low)
Contamination Source: Is it From Bugs or Dirty Dishes?
Just as a table or chair needs all of its legs to be stable, so too, all “legs of the table” must be present for a “BOD” to be determined:
You must have bugs, oxygen, and a food source
When it comes to contamination, you can have bacterial contamination, general contamination (food source) or BOTH. The key to successful troubleshooting is to identify the signs which point to whether the problem is related to general contamination, or whether the contamination is due to "bugs".
Example: GGA Fails High but Blanks are Perfect
- The contamination could be “bugs”.
The source could be a bad filter in the DI system (bugs growing within the resins).
Blanks are likely fine because glassware is clean and there is no “food source” to keep bugs going and expending oxygen.
GGA fails high due to the extra oxygen consumed by the bugs as they attack the GGA.
- The contamination could also be just “dirty glassware”.
The contamination could also be just “dirty glassware”, providing a food source.
Blanks are likely fine because –despite availability of a food source (the “crud”)--- there is no source of bugs and therefore no oxygen can be used.
GGA fails high due to the extra oxygen consumed by the bugs as they attack both the GGA and the “crud”
Example: GGA & Blanks BOTH Fail High (Blanks deplete 1-2 mg/L)
- he contamination is likely a combination of “dirty glassware”, dilution water, and “bugs”.
Blank(s) and GGA fails high because not only is there a food source (“crud”) but also there are “bugs” that shouldn’t be there.
Back to Top
Excessive DO depletion in in BOD blanks is one of the most common problems that plague all wastewater testing laboratories. Even the best labs occasionally have problems meeting the DO depletion in the BOD blanks (< 0.2 mg/L). The following are the most common causes for excessive depletion in blanks:
- Tubing is constructed of oxygen-demand leaching material
- Correct tubing not being used throughout the lab
- Slime growth in delivery tubing
- Poor water quality/improperly maintained system
- Poorly cleaned BOD bottles or dilution water storage container.
NOTE: Glass is best!
- Contamination during aeration
- Poorly calibrated DO Probe
Back to Top
Avoid “grocery store” distilled water.
This water is stored in plastic bottles which can leach oxygen demanding materials. Some analysts have had great luck using “grocery store” distilled water. If you have good luck with a particular brand, don’t change. However, be aware that you can’t control how the water is stored. If the water sits on a shelf for several months in a hot warehouse, it’s going to be a problem. Nothing leaches organic matter from plastic like hot water. Try doing a blank using dilution water prepared from a jug of distilled water that's been in the back sesat of your vehicle for a week during a particulary hot July!
Those that have successfully used “grocery store” generally have a system. Some will go directly to the factory and buy fresh water. They then date the water, store it in a cool, dry place and discard the water after it is a certain age (e.g., expiration date).
Bottom Line: Use it if it's working for you; but if it isn't try considering alternatives.
Aging Dilution Water
You should not have to age dilution water --as Standard Methods suggests-- if the water is prepared properly. If you have to age to improve quality, you should be concentrating your efforts on improving the preparation process.
Deodorizers and Dilution Water
Avoid using auto-dispensing deodorizers in the laboratory. You may be tempted to use deodorizers because wastewater labs can have a bit of an odor. However, don’t use them! They typically use alcohol as a carrier which has a very high BOD.
Back to Top
Dilution water- simplest solutions
There are a number of simple solutions to obtaining quality lab reagent water to prepare BOD dilution water.
- Obtain water from another laboratory or vendor.
Probably the easiest approach is to obtain water from another laboratory that has a track record of producing
consistently high quality BOD dilution water. It may be easier to haul water once a week than maintaining a lab
water system, particularly if you have a small laboratory.
- Purchase water from a source that has proven success.
You may also find a commercial source of water that is of consistent quality. If you find a particular brand
and supplier that works for you, .stick with it! Don’t fix it if it isn’t broke.
- Buy an all glass laboratory still and distill your own water.
If you only need a limited amount of water (e.g., < 5 gallons per week), consider buying an all glass still.
They are less costly than other water purification systems and tend to produce quality reagent water providing they are maintained properly.
- Buy a bench-top water RO and polisher combo that will produce ASTM Type I water.
Bench-top reverse osmosis (RO)/polisher combination systems that will produce ASTM Type 1 water are also available.
These systems work well. However, they tend to be expensive (> $1,000) and require regular maintenance to be effective.
Back to Top
Delivery tubing & disinfection
Disinfect delivery tube weekly, using
- dilute bleach solution: 50mL bleach per 2L of lab reagent water, or
- dilute solution of hydrochloric acid (HCl): 100 mL HCl per 1L lab reagent water
While hydrochloric acid (HCl)can be used for cleaning, it's safer to use a dilute household bleach solution.
NOTE:
- DO NOT mix acid with bleach! Chlorine gas is produced in this reaction. Even in small quantities, exposure to chlorine gas can be fatal.
- Use reinforced nylon tape around larger bottles for safety
- Nothing touches water except teflon or glass
Back to Top
Solving: Glassware cleanliness problems
- Use a good lab-grade non-phosphate detergent and bleach
- Rinse thoroughly with tap water followed by distilled water
- Allow to dry before storing.
- Always cover glassware and store in a clean, dry place.
- Commercial lab dishwashers with non-phosphate, lab grade detergents and acid rinse solutions (non-phospate) will also work well.
- Use BOD QC samples (blanks, duplicates and GGA) to evaluate washing effectiveness.
Back to Top
Solving: Water System issues
Follow manufacturer's recommendations for cleaning and disinfecting stills, etc
Simple deionizer systems can work well but can quickly be overgrown with bacteria and mold.
Can leach organics if not maintained regularly.
If using simple deionizer system, use nuclear-grade or virgin resin.
i.e., Lower grade or “re-used” resins WILL leach organic matter and cause problems.
Caution: Charcoal can become contaminated with bacteria and cause problems as well (at least one lab’s experience”).
Back to Top
Troubleshooting: Consistent high bias in GGA
Seed source selection is critical; if recycling final into primary clarifiers, could be adding nitrifiers to the seed
To determine if nitrification is occurring, try adding a nitrification inhibitor.
Compare GGAs seeded with domestic wastewater vs. commercial (Polyseed, BOD seed)
If you don’t warm the GGA before use, results will be consistently high
If nitrification is occurring:
Select another source (that does not receive final wastewater)
Use commercial seed
Back to Top
Troubleshooting: Consistent low results for GGA
- Not enough seed -
adjust the amount used until you
consistently achieve GGA results in the acceptable range.
- Poor seed quality -
try another seed source(mixed liquor; primary; another WWTP; commercially prepared seed) Recall the previous slide showing 2 vs. 4 mL of seed
- GGA too old or contaminated -
discard expired or
contaminated solutions
- Try another GGA source -
Several different types / vendors (NCL, Fisher, other scientific specialty companies)
Back to Top
Troubleshooting:Poor Precision (samples)
Poor sample precision is best characterized by wide variation among dilutions. You have to remeber, however, that BOD is a bioassay technique. This means:
- BOD is inherently less precise than instrumental tests like ammonia and total phosphorus.
[Don't expect too much]
- Look into sample measuring technique [Is the procedure used to pour sample and replicate causing variablity?]
- Look for “chunks” that might still be visible. One chunk--even a bug or piece of algae-- that goes into the sample but a similar chunk does not go into the replicate, can cause replicates to fail [ Look closely at the filter residue for the sample and replicate ]
- Are you using the appropriate precision measurement tool? [ Range works well for influents, but RPD is best for clean effluents ]
- More concern with poor precision in final vs. raw [There shouldn't BE things like "chunks" in effluent!]
Back to Top